Microscopy inFour Dimensions

An interactive look at light-sheet microscopes.

Words by  Smrithi Sunil

Designed by  Leo Laguna

Share this article

There is a microscope capable of imaging everything from the flickering dynamics of single molecules to the movements of an entire, growing embryo. Called a lattice light-sheet microscope and first developed by Eric Betzig’s group in 2014, it uses thin sheets of patterned light to capture high-resolution, three-dimensional images of subcellular structures across cells and tissues in their native environment, in seconds.

It can record transcription factors binding DNA, immune cells migrating and chromosomes separating within the cells of a growing embryo. A single instrument, in other words, can record objects spanning four orders of magnitude in both time and space.

In microscopy, there has historically been a trade-off between speed and spatial resolution.

A microscope designed to capture many images per second often has poor spatial resolution; the images appear blurry.

Wide-field fluorescence microscopy · Source ↗

A widefield fluorescence microscope can capture individual neurons firing, for example, but cannot record the molecules within those neurons.

A microscope capable of imaging at high spatial resolution, on the other hand, is often tediously slow.

Electron microscopes can resolve the structure of individual ion channels that partake in action potentials, but only after the molecules have been frozen in place; the images are static.

Light-sheet microscopes dismantled this divide. They made it possible, for the first time, for biologists to capture emergent dynamics across scales, from the stochasticity of individual molecules to the near-deterministic behavior of whole organisms. But they are also expensive and complex to operate, and so they haven’t spread to most biology labs.

For all their complexity, these instruments rely on principles that have been technologically available for about a century. No new optics were required to invent them. The great breakthrough was simply that researchers changed where they placed the light source: Imagine passing a thin sheet of light through a growing embryo from one side. If a camera is placed overhead, facing the broad side of the sheet, it can photograph a whole layer of cells in a single exposure. As the embryo moves through the sheet, successive layers of cells slide into view, quickly enough to record the embryo in three dimensions and then to watch its cells move from one image to the next.

But don’t take
our word for it.

In this article, you can deconstruct a light-sheet microscope for yourself to see how its components work together.

The Problem
With Lenses

Sheets of
Light

Ernst Stelzer’s lab at the European Molecular Biology Laboratory in Heidelberg, Germany, solved the third problem — the stretched focal length — first. Their key insight was to arrange two objective lenses perpendicular to each other; one for illumination and the other for collecting the emitted light. In 1994, they build this device (called the confocal theta microscope), showing that the elongated focal volumes fr0m each lens intersected at right angles, thus creating a more spherical volume from which fluorescence is detected.

This did not solve the speed problem, however, because the spot still had to be scanned point by point in three dimensions to build an image of the sample. About a decade later, Jan Huisken, a PhD student in Stelzer’s lab, took on the speed problem. Interested in embryonic development, they wanted to watch individual cells as they grew and divided to form fully functioning organisms. That meant capturing images sharp enough to distinguish single cells, fast enough to catch them dividing. Scanning point by point would never be fast enough, and so they decided to light up a whole plane at once.

The tool for that was a cylindrical lens. A typical convex lens curves in two dimensions, bending light into a cone that converges at a single point.

A cylindrical lens, however, curves in only one dimension, so it bends light into a wedge that narrows toward a line rather than a point.

Near that line, the wedge becomes a thin sheet of light that can illuminate an entire plane of the sample at once.

So Huisken’s team took the perpendicular objectives of confocal theta microscopy, but added a cylindrical lens. The sheet of light then entered the sample from one side, while a conventional objective, positioned at 90 degrees, collected the fluorescence from the plane the sheet lit up. The team called this selective plane illumination microscopy (SPIM).

This design addressed all three problems. The light sheet excites fluorescence in the sample, and the detection optics focus this fluorescence onto a camera.

Detection works much as it does in a conventional widefield microscope, with a camera capturing an entire plane in a single exposure, so it is just as fast. But unlike in a widefield microscope, where fluorescence from above and below the focal plane is also captured, the light sheet excites only a thin slice of the sample, greatly reducing the blurry background.

And because only that slice is lit, the thickness of the sheet, rather than the stretched focal spot of the detection lens, determines how well the microscope sees in depth.

Finally, moving the sample through the light sheet in fixed steps captures successive planes, which can be stacked to reconstruct a three-dimensional volume. This method doesn’t capture the whole volume instantly, but it only requires scanning in one dimension rather than three, which saves a lot of time.

With this system, Huisken studied embryonic development in medaka fish tagged with green fluorescent protein and in opaque fruit flies. Over the next few years, light-sheet microscopy was used to track a zebrafish from a single fertilized egg to more than 20,000 cells over its first 24 hours, following the fate of every cell.

Pushing
the Limits

Scientists now had a microscope able to image entire 3D volumes at single-cell resolution in a matter of seconds, with minimal harm to the sample. But it still could not see inside individual cells, and the reason was the sheet itself. The thickness of the sheet sets how well a light-sheet microscope sees in depth, so a sharper view requires a thinner sheet. Yet a light sheet stays thin over only a limited distance before diffraction causes it to spread out, and the thinner the sheet at its focus, the faster it spreads. Where the sheet spreads, it excites out-of-focus light that degrades the image.

This created a new trade-off, between field of view and resolution. SPIM’s 1.5-millimeter field of view had an axial resolution of 6 micrometers — fine for watching cells divide, but an order of magnitude too coarse to see anything happening inside of them.

To look inside cells across a useful field of view, then, researchers needed a sheet that was very thin and stayed thin over a long distance.

The answer, again, came from lenses. Recall that two rays from opposite edges of a lens remain in phase not just at the focus but also above and below it, along the optical axis. It is the rays from all the other parts of the lens, which travel shorter or longer distances, that fall out of phase and dim the light along the axis. What if the illumination included only the rays that travel the same distance, and all the other light were somehow blocked?

A 2011 paper from Eric Betzig’s lab did just that, shining a ring of light, rather than a full circular beam, onto the back of the objective lens. This creates what is called a Bessel beam, a narrow, pencil-like beam of light that maintains its width over a much longer distance because its waves stay in phase. A pencil beam is not a sheet, so the researchers rapidly swept it from side to side, creating a light sheet about 1 micrometer thick that stayed uniformly thin over longer distances.

But the Bessel beam brought back an old problem: blur. Light from different parts of the ring approaches the center of the beam from different directions. At the center, the waves arrive in phase and interfere constructively, creating a bright core. A little to the side, they interfere destructively, and farther still they reinforce each other again, creating bright rings called side lobes. In three dimensions, picture a bright central cylinder of light surrounded by progressively larger and dimmer hollow cylinders.

Seen from the side by the detection objective, those hollow cylinders extend above and below the plane being imaged. So the side lobes excite fluorescent molecules outside that plane, generating the same out-of-focus background that light sheets were meant to eliminate.

The Betzig lab, in 2014, found a way around this. Instead of illuminating an entire ring at the back of the objective, the researchers illuminated only parts of it, which changed where the waves would interfere constructively and destructively. The result was a 2D array of bright, pencil-like beams, called an optical lattice, in which the beams lying in the detection focal plane were much brighter than those above and below it. By rapidly moving this lattice sideways, the researchers blurred the bright beams in the focal plane into a thin, uniform sheet of light with minimal illumination above and below it.

This, finally, is the modern lattice light-sheet microscope. It achieved a lateral resolution of 150 nanometers and an axial resolution of 280 nanometers. Its depth resolution was now about 20 times finer than SPIM’s and within a factor of two of its side-to-side resolution, sharp enough to watch the inner workings of whole cells.

But studying individual cells or small groups of cells is not enough to understand how they work together to form tissues and organs. In thicker samples, such as whole tissues or small organisms, light is scattered and distorted as it passes through, degrading the image.

To correct for this, researchers borrowed adaptive optics, a technique originally developed for astronomy, which measures and corrects distortions in the light as it travels through tissue. Other add-ons pushed the resolution even further. Structured illumination, for example, combines images acquired under different illumination patterns with computational techniques to extract finer details.

Together, these advances have made it possible to observe biological processes across a large range of spatial and temporal scales, from the movement of individual molecules inside cells to the coordinated activity of entire tissues and organs. Scientists can now watch immune cells migrating through large tissues and observe how mechanical forces propagate in three dimensions.

In vivo adaptive optical super-resolution imaging comparing uncorrected tissue scattering against adaptive optics correction and 3D subcellular reconstructions
In vivo adaptive optical super-resolution imaging with MOSAIC · Fu et al., 2026 ↗

Bottlenecks
to Scaling

Each technical advance in microscopes meant more engineering, which means more specialized expertise. A talented graduate student can build an original light-sheet microscope for a few tens of thousands of dollars, but a lattice light-sheet microscope with structured illumination and adaptive optics can cost close to $1 million and needs a highly trained optical engineer to build and maintain. Efforts to make these systems more accessible have had limited success.

Light-sheet microscopy has also redefined what big data means in biological imaging. A single embryogenesis experiment that uses light-sheet microscopy can generate over 100 terabytes of data in 24 hours. That’s over a gigabyte per second. The classic setup — a personal computer sitting next to the microscope — no longer works. Labs need dedicated infrastructure to transfer, store and process the data.

Even after the images have been processed into three-dimensional volumes, extracting meaningful biological information from them is much harder. Following thousands of cells over time, identifying rare biological events or even measuring how structures change over time requires computational expertise that many biology labs do not have. The result is a gap between what the microscopes can capture and what scientists can analyze.

In addition to expertise and data, there’s also a throughput bottleneck. Researchers can’t visualize thousands of different types of molecules moving through cells and tissues at once. Fluorescent tags typically limit them to following three molecules at a time, and each tag has to be engineered and tested, which can take months, before data collection can even begin.

Just getting started with cutting-edge instrumentation, data infrastructure and expert staff would cost millions of dollars, resources typically not available to any individual biology lab.

Eric Betzig and Gokul Upadhyayula, both at the University of California, Berkeley, are trying to solve these problems by building a major microscopy platform called the Cell Observatory. They’ve equipped a room at the school with microscopes that combine lattice light-sheet imaging, structured illumination and adaptive optics with super-resolution and two-photon microscopy. By imaging the same biological sample across scales, from individual molecules to cells, tissues and whole organisms, the researchers aim to connect processes that are usually studied separately.

Ultimately, the Cell Observatory aims to use these datasets to train a foundation model capable of interpreting biological dynamics across five or more dimensions: three in space, one in time and at least one molecular marker. The microscope, however, is only one part of the larger initiative. The team is also developing dozens of fluorescent reagents for molecules of interest, building a multimillion-dollar computing facility that can process petabytes of data a day in real time and, crucially, building the model itself. The platform is built to handle multiple organisms, including zebrafish, mouse embryos and even human organoids.

Multimodal 3D fluorescence imaging of cellular and subcellular dynamics inside a living zebrafish embryo with segmented individual cells
Cellular and subcellular dynamics in a living zebrafish embryo (MOSAIC) · Fu et al., 2026 ↗

Biology is not static.

Molecules are constantly moving, cells are dividing and responding to other cells, and organs are performing specialized functions. For most of the history of microscopy, improving our ability to see smaller structures meant sacrificing our ability to watch them move. While light-sheet microscopy does not eliminate this trade-off completely, it has dramatically expanded the spatiotemporal range of biological processes we can observe in an organism in its natural environment. The challenge now is to put these capabilities into the hands of more biologists.


Smrithi Sunil is a biomedical scientist and science writer who has invented microscopy techniques to study the brain. She holds a doctorate in biomedical engineering from Boston University.

Notes

  1. Picture light traveling from two opposite ends of the lens to form a triangle at the focus. A ray traveling at the center of the lens would split that triangle into two right triangles. So the distance traveled by the center ray and the edge ray is similar to the adjacent side and the hypotenuse of a right triangle.
  2. When light waves interfere, the resulting amplitude is the sum of amplitudes of all the waves combined. Two waves of equal amplitude that are perfectly in phase produce a wave with twice the amplitude, while two waves that are perfectly out of phase cancel each other completely.
  3. This point will come in handy later in the essay.
  4. Another advantage of a thin light-sheet is the reduction in phototoxicity, and therefore interference with natural biological processes. While confocal illuminates the entire depth of its illumination cone at every scanned point even though it only detects light from the focal point, exposing far more of the specimen to damaging light than it ever images, light-sheet illuminates only the thin plane being imaged at any moment.